Adjust actuator range, taking into account the measured range of motion
This follows on from !512 (merged) and !513 (merged), and the measurements from #460 (closed) :
A new set of measurements examined the actuator travel by counting turns of the actuator.
- There were ~15.5 turns of the M3 screw from when the stage started to move as the actuator lifted off the foot, to when the actuator stopped moving at the top. The central position was at ~7.75 turns, which is the mid point. This is about 7.8mm travel, which is between 7.2mm and 8.2mm as expected, and indicates that ~0.4mm of the allowed clearance is actually used by printer artifacts.
- A linear regression of the measured stage positions after each turn of the M3 screw showed 0.762+-0.001mm per turn. With 0.5mm pitch of the screw, that gives a lever ratio of 1.52.
The total range of actuator motion in turns is about 7.8mm travel, which indicates that ~0.4mm of the allowed clearance is actually used by printer artifacts (design motion including clearance was 8.2mm).
- This MR sets the clearance to
0.2mm each endNote, in the final version it is 0.4mm top and 0.2mm bottom.
The stage xy motion is a lot less than expected - there is a reduced effective lever ratio compared with the 7/4 design ratio
- This MR introduces an extra parameter to take into account the measured effective lever ratio, which is used as an additional adjustment to the 'nubs' on the actuator columns, and to the position of the end stops in the feet.
Additional information from the discussion and comments in this MR:
There is extra information on the effective lever ratio from two papers which broadly agrees with my measurements above.
- Sharkey et al. 2016 https://doi.org/10.1063/1.4941068 found that the motion was ~86% of the expected travel per step (lever ratio 1.50). This is on a much older version of the stage, but with the same xy lever geometry.
- Chisholm et al. 2026 https://doi.org/10.1111/jmi.70192 found that the effective lever ratio was 1.54 on a recent v7 beta body.
So an effective ratio of 1.52 seems to be appropriate.
- Testing of a couple of main bodies in different places shows that the end stops do not give an entirely reproducible range of motion, and that the total range in y might be less than in x. However this MR does give a range which is close to the expected range without allowing over-flexing of the actuators.
- Range of motion achieved is not above 13mm sometimes seems to be less than 12mm.
closes #460 (closed) Issue #489 has been made to continue discussion.
Notes on measurement fixtures:
There is a SCAD file in accessories which provides fixtures to allow simple measurement of the motion of the stage top using a dial gauge.
This is not designed for use with the microscope stand, I have been using it with manual operation on a bare microscope body. This allows me to move quickly from one end of motion to the other.