Clarify objective specs with inclusion of conjugate distance.

This MR:

  • Further highlights need for objective to be Plan corrected
  • Mentions that the parfocal length is sometime specified as a conjugate distance (195mm for 45mm parfocal length)
  • Highlights that 185mm conjugate distance objectives (often with just 185 in product description) are not the correct objective for the standard optics module.

Closes #377 (closed)