[TESTING] - Imaging | Fluorescent agent

Description

Test the imaging system's ability (with the excitation LED module) to capture fluorescence levels of a fluorescent agent, perfused in a microfluidic chip. In other words, create a standard/calibration curve.

image

Notes:

  • The agent is fluorescein sodium or any other dye close to GFP's excitation/emission spectrum. You can also try to use its pH dependent fluorescence properties, as explained here.
  • Check out this protocol on a fluorescence standard curve with fluorescein in Calibration 3: Fluorescence standard curve - Fluorescein Protocol
  • To simulate different levels, you can perform serial dilutions of the agent in water or PBS
  • Single-layer chips can be used
  • Most likely, you won't saturate the signal so your curve, well, will be a linear line.
Edited by Asterios Arampatzis